深圳市宏美瑞科技有限公司為您介紹美國泰克電流探頭P6021A:
泰克電流探頭P6021A and P6022 Current Probes provide versatile AC current measurements. Both probes provide accurate current measurements over a wide range of frequencies. The P6021 and P6022 allow current measurements without breaking the circuit by clipping onto the current carrying conductor. Shielded probe heads are not grounded when the slides are in their open positions, eliminating accidental grounding of the circuit under test.
KEY PERFORMANCE SPECIFICATIONS
- 120 Hz to 60 MHz
- 10.6 A RMS, 250 A peak, 10 mA sensitivity
- P6022
- 935 Hz to 120 MHz
- 4 A RMS, 100 A peak, 1 mA sensitivity
KEY FEATURES
- For 1 MΩ inputs
- Shielded probe head
- AC only
- Split core construction allows easy circuit connection
- 1.5 m (5 ft) cable
APPLICATIONS
- Motor drives
- Power inverters/converters
- Power supplies
- Avionics
For general purpose applications, the P6021A provides wide-band performance with excellent low-frequency characteristics. Bandwidth is 120 Hz to 60 MHz. The probe range is switchable between 2 mA/mV and 10 mA/mV.
P6022
With a head size of 0.47 in. x 0.25 in. (10 mm x 6 mm, about half the size of the P6021A) and a bandwidth of 935 Hz to 120 MHz, the P6022 is ideal for measuring currents in compact, high-performance circuits. Passive termination output is switchable between 1 mA/mV and 10 mA/mV.
Cable length1.5 m (59 in)
P6021A probe headLength20 cm (7.77 in)
Width16 mm (0.625 in)Height32 mm (1.25 in)
Maximum conductor diameter5 mm (0.197 in)
P6022 probe headLength152 mm (6.0 in)Width6.4 mm (0.25 in)
Height12 mm (0.47 in)
Maximum conductor diameter2.8 mm (0.11 in)
EMC ENVIRONMENT AND SAFETY
Compliance
CAN/CSA-C22.2 No. 61010-1
CAN/CSA-C22.2 No. 61010-2-032
UL 61010-1
UL61010B-2-032
EN 61010-1
EN 61010-2-032
Ordering information
MODELS
P6021ACurrent ProbeP6022Current Probe with termination
STANDARD ACCESSORIES
6 in. ground lead196-3521-00 Instruction manual071-3004-00 (P6021A),070-0948-03 (P6022)Termination011-0106-00 (P6022 only)
RECOMMENDED ACCESSORIES
Nylon carrying case016-1952-xxCurrent loop, 1 turn, 50 Ω with BNC connector, used for Performance Verification067-2396-xxDeskew/calibration fixture067-1686-xx
WARRANTY
One year parts and labor.
Service options
Opt. C3Calibration Service 3 YearsOpt. C5Calibration Service 5 YearsOpt. D1Calibration Data ReportOpt. D3Calibration Data Report 3 Years (with Opt. C3)Opt. D5Calibration Data Report 5 Years (with Opt. C5)Opt. R3Repair Service 3 Years (including warranty)Opt. R3DW
Repair Service Coverage 3 Years (includes product warranty period). 3-year period starts at time of instrument purchase
Opt. R5Repair Service 5 Years (including warranty)Opt. R5DW